Introduction

Percutaneous coronary interventions (PCI) in bifurcations are still challenging and are associated with higher risks of periprocedural complications as well as in-stent restenosis and stent thrombosis [1, 2]. Dedicated bifurcation stents were proposed as one of the resolutions to these issues [3]. However, few of them have been tested in large randomized clinical trials [4, 5].

In response to this interest, the Polish dedicated bifurcation stent was designed. The BiOSS Clinical Program started in 2008. Initially, the BiOSS stent was the only stainless steel stent, but shortly after in 2010 a paclitaxel-eluting stainless steel stent, the BiOSS Expert stent, was developed. Moreover, going with the times, in 2012 the sirolimus-eluting stainless steel BiOSS LIM stent was released. We gathered a lot of data coming from both randomized clinical trials as well as from real world evidence (RWE) studies [614]. The data were acceptable and encouraging, but still a way for enhancement was sought. In consequence, a cobalt-chromium sirolimus-eluting version was created – the BiOSS LIM C stent. Research on an animal model met the expectations as to the safety profile and effectiveness, and thus the BiOSS LIM C stent started the phase of clinical trials [15].

Previously, we presented 3-month results of the interim analysis of the BiOSS LIM C Registry [16]. The aim of this research was to summarize 12 months’ complete clinical results of the prospective, first-in-man BiOSS LIM C registry.

Material and methods

Study population and device description

Between August 2016 and December 2017 we enrolled consecutive patients meeting exclusion and inclusion criteria with stable coronary artery disease or non-ST-elevation acute coronary syndrome (NSTE-ACS). Procedures were performed in two centers in Poland (Warsaw, Olsztyn). Prior to the procedure, written, informed consent was obtained. The Local Ethics Committee approved the registry protocol (No. 84/2016).

The BiOSS LIM C is a balloon-expandable sirolimus-eluting stent made of cobalt-chromium (strut thickness 70 μm). Sirolimus (1.4 μg/mm2) is released from the biodegradable coating built of a copolymer of lactic and glycolic acids (Figure 1). The BiOSS LIM C stent can be characterized as having two parts different in the diameter. The proximal part is joined with the distal part by only two struts. The details of patient population/device description were presented in detail previously [16].

Figure 1

A – BiOSS LIM C stent alone, B – bottle balloon catheter, C – BiOSS LIM C crimped on the balloon, D – the fully deployed BiOSS LIM C

https://www.archivesofmedicalscience.com/f/fulltexts/111580/AMS-19-2-111580-g001_min.jpg

Interventional procedure and concomitant medications

Provisional T-stenting (PTS), i.e. single stent deployment in the main vessel-main branch across a side branch, was the default strategy in all patients. Medina classification was used to assess bifurcation lesions, where the value of 1 indicated stenosis greater than 50% (visual estimation) [17]. No restriction was applied in terms of lesion length. If necessary, an additional regular sirolimus-eluting stent was deployed. A drug-eluting stent was deployed in a side branch only if proximal residual stenosis was greater than 70% after balloon dilatation and/or significant blood flow disturbances after main vessel–main branch stenting and/or blood flow limiting dissection were observed.

The stent deployment protocol was based on the recent European Bifurcation Club guidelines and it was as follows [18]:

  • wiring of both branches,

  • main vessel predilatation and/or side branch predilatation according to the operator’s judgment,

  • stent deployment (inflation for at least 20 s),

  • proximal optimization technique (POT; highly recommended, but at operator’s judgment),

  • side branch postdilatation/side branch stent deployment if required,

  • final kissing balloon inflation after 2nd stent implantation or at operator’s judgment in other cases,

  • re-POT after final kissing balloon technique.

Proximal optimization technique is a technique in which a short balloon is expanded in the proximal part of the stent in the main vessel. This ensures full expansion and complete apposition of the stent as well as facilitating re-crossing into the side branch. Final kissing balloon technique is mandatory when two stents are implanted, and is based on simultaneous inflation of two balloons – one located in the main vessel/main branch, and one located in the side branch.

In patients with NSTE-ACS a loading dose of P2Y12 inhibitor was given (clopidogrel 600 mg or ticagrelor 180 mg), and if necessary, also a loading dose of acetylsalicylic acid (ASA) was administered (300 mg). In planned cases 72 h before the procedure patients received ASA (75 mg/day) and clopidogrel (75 mg/day) or ticagrelor (90 mg b.i.d). All percutaneous coronary interventions (PCIs) were performed via radial or femoral access using 6 Fr or 7 Fr guiding catheters, and periprocedurally unfractionated heparin was administered (100 IU/kg). Dual antiplatelet therapy was recommended for 12 months.

All patients had TnI, CK and CK-MB levels determined before PCI, and 6 and 24 h after. Periprocedural myocardial infarction (MI) (type 4a) was defined according to the third universal definition [19].

Endpoints

The composite primary endpoint (major adverse cardiovascular events – MACE) was defined as the rate of cardiac death, MI and clinically driven target lesion revascularization (TLR) in 12-month follow-up. Secondary endpoints were as follows: death rate, MI rate, stent thrombosis (ST) rate as well as the device, angiographic and procedure success. Clinically driven TLR was defined as the reintervention within the target lesion because of a symptomatic ≥ 50% diameter stenosis within 12 months after the index procedure. Further details were presented previously [16].

Statistical analysis

Continuous variables are shown as mean ± standard deviation. Categorical data are shown as numbers (%). Continuous variables were compared using an unpaired two-sided Student t-test, and categorical data using the χ2 test or Fisher’s exact test, as appropriate. If the distribution was not normal in the Shapiro-Wilk test, the Wilcoxon signed-rank and Mann-Whitney U-tests were applied. P-values of < 0.05 were recognized as statistically significant. The time to event analysis was performed with the Kaplan-Meier estimator of survival curves. Statistical analyses were performed using R 3.0.2 software for OS (R Foundation, Vienna, Austria).

Results

The study population consisted of 95 patients. The mean age was 66.8 ±9.8 years and 17.9% (n = 17) were females. In the study population hypertension (90.5%) and dyslipidemia (81.1%) were the most common risk factors. The target lesion was located in the left main most commonly (55.8%, n = 53) (Table I). In Table II baseline clinical characteristics in subgroups LM vs. non-LM are presented. In the LM subgroup significantly more patients had diabetes (43.4% vs. 21.4%, p < 0.05) and dyslipidemia (84.9% vs. 52.4%, p < 0.05). Baseline angiographic characteristics are presented in Table III. It is noteworthy that true bifurcations were present in 52.6% (n = 50).

Table I

Baseline characteristics

ParameterBiOSS group, n = 95 (%)
Age [years]66.8 ±9.8
Women (%)17 (17.9)
Hypertension86 (90.5)
Hypercholesterolemia77 (81.1)
Diabetes type 232 (33.7)
Prior MI51 (53.7)
Prior PCI44 (46.3)
CABG14 (14.7)
Peripheral artery disease10 (10.5)
Chronic kidney disease13 (13.7)
History of smoking21 (22.1)
Clinical indication for PCI:
 Planned PCI70 (73.7)
 UA14 (14.7)
 NSTEMI10 (10.5)
 STEMI0
Lesion location:
 LM53 (55.8)
 LAD26 (27.4)
 LCx13 (13.7)
 RCA3 (3.2)

[i] MI – myocardial infarction, PCI – percutaneous coronary intervention, CABG – coronary artery bypass graft, UA – unstable angina, NSTEMI – non-ST elevation myocardial infarction.

Table II

Baseline characteristics in left main (LM) and non-LM subgroups

ParameterLM subgroup, n = 53 (%)Non-LM subgroup, n = 42 (%)
Age [years]68.3 ±8.665.8 ±7.4
Women (%)12 (22.6)5 (11.9)
Hypertension47 (88.7)39 (92.9)
Hypercholesterolemia45 (84.9)22 (52.4)*
Diabetes type 223 (43.4)9 (21.4)*
Prior MI30 (56.6)21 (50)
Prior PCI28 (52.8)16 (38.1)
CABG11 (20.8)3 (7.1)
Peripheral artery disease6 (11.3)4 (9.5)
Chronic kidney disease6 (11.3)7 (16.7)
History of smoking13 (24.5)8 (19.1)

MI – myocardial infarction, PCI – percutaneous coronary intervention, CABG – coronary artery bypass graft,

* p < 0.05 between LM and non-LM groups.

Table III

Baseline angiographic characteristics

ParameterWhole population, n = 95 (%)LM subgroup, n = 53 (%)Non-LM subgroup, n = 42 (%)
SYNTAX score26.09 ±3.427.6 ±4.224.2 ±2.8
EuroSCORE II0.86 ±0.210.94 ±0.320.76 ±0.29
Medina classification:
 1.1.122 (23.2)12 (22.6)10 (23.8)
 1.0.113 (13.7)6 (11.3)7 (16.7)
 0.1.115 (15.8)7 (13.2)8 (19.1)
 1.0.012 (12.6)8 (15.1)4 (9.5)
 0.1.010 (10.5)6 (11.3)4 (9.5)
 1.1.023 (24.2)7 (13.2)16 (38.1)

[i] LM – left main.

The procedure characteristics are shown in Table IV. The device and angiographic success rates were 100%. The procedure success rate was 95.8%. Main branch predilatation was performed in 80% of cases. The side branch required additional dilatation in over a third of lesions (38.9%), and in 18 cases (18.9%) the classical DES was deployed, mostly in T and protrusion (TAP) technique (n = 15, 15.8%). The proximal optimization technique was performed in 53.7% of cases, and the final kissing balloon inflation was used in 29 (30.5%) patients.

Table IV

Procedural characteristics

ParameterBiOSS group N (%)LM subgroup n = 53 (%)Non-LM subgroup n = 42 (%)
Successful implantation95 (100)53 (100)42 (100)
Main vessel predilatation76 (80.0)45 (84.9)31 (73.8)
Side branch predilatation25 (26.3)20 (37.7)5 (11.9)*
Both branches’ predilatation19 (20.0)17 (32.1)2 (4.8)*
Nominal stent diameter in main vessel [mm]3.89 ±0.454.17 ±0.363.78 ±0.35*
Nominal stent diameter in main branch [mm]3.18 ±0.423.42 ±0.353.01 ±0.23*
Nominal stent length [mm]19.51 ±3.2919.94 ±3.1618.78 ±3.43
Side branch postdilatation37 (38.9)23 (43.4)14 (33.3)
POT51 (53.7)37 (69.8)14 (33.3)*
FKB29 (30.5)22 (41.5)7 (16.7)*
Additional stent in SB18 (18.9)15 (28.3)3 (7.1)*
Dissection requiring an additional stent in main vessel11 (11.6)7 (13.2)4 (9.5)
Fluoroscopy time [min]19.6 ±12.421.2 ±12.117.6 ±9.5
Contrast volume [ml]197 ±66211 ±75180 ±43
Vascular access femoral/radial6.3%/93.7%5.7%/94.3%7.1%/92.9%
Guiding catheter 6F/7F95.8%/5.3%90.6%/9.4%100%/0

LM – left main, POT – proximal optimization technique, FKB – final kissing balloon;

* p < 0.05 between LM and non-LM groups.

Myocardial infarction type 4a was registered in 4 (4.2%) patients and was caused by temporary side branch occlusion. In the 12-month follow-up the MACE rate was 9.5% (n = 9), including 1 (1.1%) cardiac death due to heart failure exacerbation, 2 (2.1%) MI and 6 (6.3%) clinically driven TLR (Table V). Kaplan-Meier curves are presented in the Figure 2. All incidents, apart from one TLR, appeared in the LM subgroup (Table VI). In the case of TLR 2 cases were treated with plain old balloon angioplasty (POBA), 3 with another DES and 1 with CABG (Figure 3).

Table V

BiOSS LIM C clinical results

Parameter1 month n = 95 (%)3 months n = 95 (%)6 months n = 95 (%)12 months n = 95 (%)
MACE01 (1.1)3 (3.2)9 (9.5)
Death001 (1.1)**2 (2.1)#
Cardiac death0001 (1.1)
MI4 (4.2)*002 (2.1)
TLR clinically driven01 (1.1)3 (3.2)6 (6.3)

MACE – major adverse cardiovascular event, MI – myocardial infarction, TLR – target lesion revascularization.

* Periprocedural MI type 4a.

** Ruptured aortic aneurysm.

# Mediastinitis.

Table VI

Clinical results of BiOSS LIM C: LM vs. non-LM

Parameter12 months n = 95 (%)LM n = 53 (%)Non-LM n = 42 (%)
MACE9 (9.5)8 (15.1)1 (2.4)
Death2 (2.1)#2 (3.8)0
Cardiac death1 (1.1)1 (1.9)0
MI2 (2.1)2 (3.8)0
TLR clinically driven6 (6.3)5 (9.4)1 (2.4)

MACE – major adverse cardiovascular event, MI – myocardial infarction, TLR – target lesion revascularization.

# Mediastinitis.

Figure 2

Kaplan-Meier curves showing event-free survival for major adverse cardiovascular events (A) and target lesion revascularization (B), respectively

https://www.archivesofmedicalscience.com/f/fulltexts/111580/AMS-19-2-111580-g002_min.jpg
Figure 3

Restenosis location in BiOSS LIM C stent

https://www.archivesofmedicalscience.com/f/fulltexts/111580/AMS-19-2-111580-g003_min.jpg

Discussion

The 12-month results of BiOSS LIM C deployment are acceptable and encouraging. The stent was successfully deployed in 100% without major difficulties, even in direct stenting. The MACE rate was 9.5%, and the clinically driven TLR rate 6.3%.

When analyzing results, a high predilatation rate was observed (80%). It could have been associated with a significant portion of true bifurcations (52.6%). However, the relatively low proportion of POT cases might be surprising. Nevertheless, one must bear in mind the POT-like effect associated with the BiOSS LIM C structure. When sizing the BiOSS stent the operator might overestimate the main vessel diameter and therefore be convinced that he obtains a POT-like effect. However, we still believe that POT should be highly recommended.

BiOSS LIM C is the improved version of BiOSS stent differing in strut thickness (70 μm vs. 120 μm). This difference might translate into better clinical outcomes. The results observed in the porcine model with the BiOSS LIM C stent seemed to confirm our hope [20]. Interestingly, Table VII presents numerical comparisons of clinical results between BiOSS LIM C, BiOSS LIM and BiOSS Expert. The results suggest that BiOSS LIM C might be superior to BiOSS Expert, but the advantages of BiOSS LIM C over BiOSS LIM need further elucidation, especially when we analyze the LM population.

Table VII

12-month follow-up from three BiOSS Registries

ParameterBiOSS LIM C, n = 95 (%)BiOSS LIM, n = 60 (%)BiOSS Expert, n = 63
MACE9 (9.5)3 (5)9 (14.3)
Death2 (2.1)1 (1.67)2 (3.2)
Cardiac death1 (1.1)1 (1.67)2 (3.2)
MI2 (2.1)1 (1.67)0
TLR clinically driven6 (6.3)1 (1.67)7 (11.1)

[i] MACE – major adverse cardiovascular event, MI – myocardial infarction, TLR – target lesion revascularization.

In the BiOSS LIM LM Registry the rate of MACE rate was 9.5% (n = 7), whereas the rate of clinically driven TLR was 6.8% (n = 5) [7]. In the BiOSS LIM C Registry, these values were 15.1% (n = 8) and 9.4% (n = 5), respectively. These results are difficult to explain since the rate of true bifurcations, nominal stent parameters, and the rate of FKB were comparable or even higher in the case of the POT rate. Nevertheless, the factors negatively affecting the BiOSS LIM C results might be as follows: SYNTAX score (27.6 ±4.2 vs. 22.42 ±4.38) and the rate of additional stent implantation in the side branch (28.3% vs. 14.9%). Also, one should remember that the difference in radial force exerted by different stent platforms (cobalt-chromium vs. stainless steel) could play a significant role [21].

However, regarding the true value of the BiOSS LIM C stent, the two ongoing multicenter international studies will decide. The first one, POLBOS 3 (NCT03548272), will assess the stent in non-LM coronary bifurcations, and the second, POLBOS LM (NCT03508219), will evaluate the use of the BiOSS LIM C stent in the treatment of LM distal stenosis.

And finally, we should put our registry into the perspective of the European Society of Cardiology guidelines on myocardial revascularization from 2018 [22]. They recommend PCI in LM in patients with a low (class I) or intermediate (class IIa) SYNTAX score, and they recommend PCI in three-vessel CAD without diabetes only in patients with a low SYNTAX score (class I). In patients with three-vessel disease, diabetes and a low SYNTAX score PCI has class IIb recommendations. In our study population 33.7% of patients had diabetes and the mean SYNTAX score was 26.09 ±3.4. When choosing between PCI and CABG it is also worth remembering about the patient’s preferences, prior history of CABG, comorbidities (e.g. severe chronic obstructive pulmonary disease) and coronary artery anatomy (especially in patients with diabetes), in which sometimes there is no possibility to implant a well-functioning graft [23, 24].

The number of patients is relatively small. Patients with coronary bifurcation lesions a priori planned for a two-stent strategy were excluded. Also, no identical deployment techniques were used. However, PCIs were performed by operators experienced in BiOSS stent deployment. And finally, no control group was gathered to compare the use of this dedicated bifurcation stent and stenting with other devices and techniques.

In conclusion, our registry might suggest that PCI using the BiOSS LIM C in coronary bifurcations is feasible and might be the option for percutaneous revascularization.