Clinical research
One-year hemodynamic comparison of Perimount Magna with St Jude Epic aortic bioprostheses
 
More details
Hide details
 
Submission date: 2011-09-27
 
 
Final revision date: 2012-01-20
 
 
Acceptance date: 2012-03-11
 
 
Online publication date: 2013-06-21
 
 
Publication date: 2013-06-01
 
 
Arch Med Sci 2013;9(3):445-451
 
KEYWORDS
TOPICS
ABSTRACT
Introduction: Cardiac surgeons are using more bioprosthetic valves due to the ageing population as well as to improvements that have been made to these implants. We sought to compare the 1-year hemodynamics of two commercially available valves by echocardiographic parameters.
Material and methods: Retrospective review of our institutional database revealed 69 patients who received either Perimount Magna (n = 33) or St Jude Epic (n = 36) valves in the aortic position with no other valve surgery between June 2004 and March 2006. All patients received transthoracic echocardiography at 1 year. Comparisons between groups were made at baseline and at 1-year follow-up. In addition, a pairwise comparison was performed in each patient to determine the change in echocardiographic parameters between baseline and follow-up.
Results: Mean implanted valve size was similar (Magna 24.3 ±2.0 mm vs. Epic 24.1 ±2.2 mm). Pre- and intraoperative patient variables were similar between the two groups. There were lower peak and mean pressure gradients in the Magna group, both at discharge and one year after surgery. This correlated with a larger indexed effective orifice area (Magna 0.8 ±0.2 cm2/m2 vs. Epic 0.67 ±0.2 cm2/m2, p = 0.02). In spite of these findings, left ventricular mass regression was not different.
Conclusions: These findings suggest that in a series with relatively low indexed effective orifice areas, the peak and mean gradients obtained were acceptable. More clinical follow-up of these patients is required to assess the true impact of prosthesis patient mismatch.