Clinical research
Ambulatory Holter monitoring in asymptomatic patients with DDD pacemakers – do we need ACC/AHA Guidelines revision?
 
More details
Hide details
 
Submission date: 2011-03-06
 
 
Acceptance date: 2011-06-19
 
 
Online publication date: 2013-11-05
 
 
Publication date: 2013-10-31
 
 
Arch Med Sci 2013;9(5):815-820
 
KEYWORDS
TOPICS
ABSTRACT
Introduction: We sought to determine the usefulness of ambulatory 24-hour Holter monitoring in detecting asymptomatic pacemaker (PM) malfunction episodes in patients with dual-chamber pacemakers whose pacing and sensing parameters were proper, as seen in routine post-implantation follow-ups.
Material and methods: Ambulatory 24-hour Holter recordings (HM) were performed in 100 patients with DDD pacemakers 1 day after the implantation. Only asymptomatic patients with proper pacing and sensing parameters (assessed on PM telemetry on the first day post-implantation) were enrolled in the study. The following parameters were assessed: failure to pace, failure to sense (both oversensing and undersensing episodes) as well as the percentage of all PM disturbances.
Results: Despite proper sensing and pacing parameters, HM revealed PM disturbances in 23 patients out of 100 (23%). Atrial undersensing episodes were found in 12 patients (p < 0.005) with totally 963 episodes and failure to capture in 1 patient (1%). T wave oversensing was the most common ventricular channel disorder (1316 episodes in 9 patients, p < 0.0005). Malfunction episodes occurred sporadically, leading to pauses of up to 1.6 s or temporary bradycardia, which were, nevertheless, not accompanied by clinical symptoms. No ventricular pacing disturbances were found.
Conclusions: Asymptomatic pacemaker dysfunction may be observed in nearly 25% of patients with proper DDD parameters after implantation. Thus, ambulatory HM during the early post-implantation period may be a useful tool to detect the need to reprogram PM parameters.